Key takeaways
- Check that the laboratory is accredited to ISO/IEC 17025 and that the specific test method falls within its accredited scope.
- A report applies only to the sample described in it, so confirm that the sample is the finished product being offered and not just the fabric.
- Separate the test method from the product standard, and make sure each result is judged against a stated requirement.
- A type-test report and a batch certificate of analysis answer different questions, and a complete file contains both.
Purchase files for drapes and gowns fill up quickly with test reports, and most of them are never read beyond the first page. That is understandable: the documents are dense, each laboratory lays them out differently, and the conclusion is often not stated in plain words. Yet a report is evidence only for what it actually says, about the sample it actually names. Reading one properly takes about ten minutes and follows the same sequence every time.
Start with the laboratory, not the result
The first question is whether the laboratory is competent to perform the test. The recognised measure is accreditation to ISO/IEC 17025, the international standard for testing and calibration laboratories. In India the accreditation body is NABL; other countries have their own bodies, linked through international mutual recognition arrangements.
Accreditation is not a blanket approval. It is granted for a defined scope: a list of specific tests, methods and product groups. A laboratory accredited for textile strength testing is not, by that fact, accredited for microbial penetration or biocompatibility. Check three things:
- The report carries the accreditation body’s symbol and the laboratory’s certificate number.
- The accreditation was valid on the date the testing was done.
- The test method shown in the report appears in the laboratory’s published scope, which can be looked up on the accreditation body’s website.
Laboratories sometimes report accredited and non-accredited tests in one document and are expected to mark which is which. A manufacturer’s own laboratory can produce perfectly sound data, but it is not independent, and its reports should be read as first-party evidence.
Identify the sample
The sample description is the most important and most neglected part of a report. It records what the laboratory received, usually in the submitter’s own words, and the results apply to that item alone. Look for:
- the product name, model or code, and whether it matches what is being offered to you;
- the material description, including construction and basis weight;
- the batch or lot number and the quantity submitted;
- the name of the party who submitted it, which may be a fabric mill and not the drape manufacturer;
- whether the sample was sterilised, and which part of the product was tested.
Most reports state that the results relate only to the sample as received, and that sampling was done by the customer. Take that statement literally. The laboratory is not vouching for production as a whole.
Roll goods or finished product
A report on fabric taken from a roll shows what the fabric can do. It says nothing about seams, lamination, the reinforced zone or the effect of sterilisation, all of which influence the performance of the finished article. For a gown, the classification depends on the weakest part of the critical zone, including the seams, as explained in our article on the critical zones of a surgical gown. Fabric data is useful, but a performance claim for a product should rest on tests of the product.
Check the method and its edition
Every result should be tied to a named test method with its year of publication. Two different matters are often run together here.
- The test method describes how a property is measured, such as a hydrostatic pressure test or a tensile test.
- The product standard sets the requirement the result has to meet, such as IS 17334:2019, EN 13795-1 or ANSI/AAMI PB70, and usually specifies which methods are to be used.
A report may quote a correct method but compare the result with a requirement from a different standard, or with no requirement at all. Methods that sound alike are not always equivalent: a result obtained by one hydrostatic test cannot simply be read across to another with different conditions. The individual tests are described in our guide to barrier tests. Note also any deviations from the method that the laboratory has recorded.
Read the results against the requirement
A number on its own means little. For each property, find the result, the unit, the number of specimens tested and the requirement it is being judged against. Useful questions are:
- Is the figure a mean, a minimum or a set of individual values? Some standards judge on individual specimens, so a satisfactory mean can hide a failure.
- Is the unit the one used in the standard? Pressure, for instance, may be reported in different units in different laboratories.
- Does the report state a pass or fail, and against which standard, class or level?
- Were all the properties required by the standard tested, or only a convenient selection?
The last point matters. A phrase such as “tested as per” a standard can be accurate when only one of its several requirements was examined. Conformity to a product standard means meeting all the applicable requirements, a distinction discussed in our comparison of IS 17334, EN 13795 and AAMI PB70.
Look at the date and the type of document
A report records one event. Materials, suppliers and processes change, and a report several years old may describe a product that is no longer made in the same way. There is no universal validity period, so tender documents often set their own; where they do not, ask whether anything has changed since the test.
It also helps to recognise which kind of document is in front of you.
| Document | Issued by | What it shows | Main limitation |
|---|---|---|---|
| Type-test report | Independent accredited laboratory | A submitted sample met stated requirements when tested | Covers that sample only, at that date |
| Raw-material report | Laboratory, for the fabric supplier | Properties of roll goods | Does not cover the finished, sterilised product |
| Certificate of analysis | Manufacturer, for each batch | Routine checks carried out on the batch supplied | First-party; limited range of properties |
| Declaration of conformity | Manufacturer | A formal claim that the product meets named standards | A claim, to be supported by reports |
| Product certification licence | Certification body | Ongoing conformity under surveillance | Applies only to the licensed product and site |
The type-test report and the certificate of analysis answer different questions. The first shows that the design is capable of meeting the standard; the second shows that a particular batch was checked. A sound file contains both.
A short checklist
- The laboratory is accredited, and this method is within its scope.
- The sample described is the product offered, in finished form.
- The method and edition match those named in the standard.
- Each result is compared with a stated requirement, with a clear outcome.
- All relevant properties are covered.
- The report is complete, with every page present, and recent enough to represent current production.
- The authenticity of the report can be confirmed with the issuing laboratory if there is doubt.
Shrisay Medical, a manufacturer of surgical drapes and gowns in Faridabad, can provide the documents relevant to a product under evaluation; details are on our quality and certifications page, or you may contact our team.
Frequently asked questions
How old can a test report be?
There is no universal limit. Many tenders set their own period, and in any case the report should represent the materials and process currently in use.
Is a NABL-accredited laboratory accredited for every test it performs?
No. Accreditation is granted for a defined scope of tests and methods. Check that the method shown in the report appears in the laboratory’s published scope.
Is a fabric test report enough to support a gown’s barrier claim?