Key takeaways

  • IS 17334, EN 13795 and ANSI/AAMI PB70 all describe barrier performance, but they are separate documents with different scopes, tests and classifications.
  • AAMI PB70 classifies liquid barrier performance in four levels, while EN 13795 uses two performance classes and also covers cleanliness, linting and strength.
  • A rating under one framework does not automatically translate into a rating under another; equivalence has to be shown by testing.
  • Specify the framework your regulator and your tender require, then state the class or level explicitly.

A buyer comparing quotations for surgical gowns may see “IS 17334 compliant” on one, “EN 13795 high performance” on another and “AAMI Level 3” on a third. All three describe barrier performance, and none of them is wrong, but they are not saying the same thing. Each comes from a different standards body, uses different test methods and classifies products in a different way. This article sets the three side by side so that claims can be compared sensibly. It works at the level of structure and intent; for limits and test conditions, refer to the current official text of each standard.

Where each standard comes from

IS 17334:2019 is published by the Bureau of Indian Standards and is the national specification for surgical gowns and surgical drapes. It is the reference most often used in Indian public procurement.

EN 13795 is a European standard covering surgical drapes, gowns and clean air suits used as medical devices. Manufacturers placing these products on the European market normally use it to demonstrate performance.

ANSI/AAMI PB70 is published by the Association for the Advancement of Medical Instrumentation and is the reference used in the United States. It covers the liquid barrier performance and classification of protective apparel and drapes intended for use in healthcare facilities, which includes isolation gowns as well as surgical gowns and drapes.

The comparison at a glance

AspectIS 17334:2019EN 13795ANSI/AAMI PB70
Issued byBureau of Indian StandardsEuropean Committee for Standardization (CEN)AAMI, as an American National Standard
Main marketIndiaEuropean Union and markets that follow European practiceUnited States and markets that follow American practice
ProductsSurgical gowns and surgical drapesSurgical gowns and drapes; clean air suits in a separate partSurgical gowns, isolation gowns, other protective apparel and drapes
ClassificationGraded by barrier performance as defined in the standardTwo classes: standard performance and high performanceFour levels: Level 1 (lowest) to Level 4 (highest)
Product zonesAs defined in the standardCritical and less critical product areasCritical zones; the product is rated by its lowest-performing critical zone component
Properties addressedBarrier and associated material properties as listed in the standardLiquid and microbial penetration, cleanliness, particle release, strengthLiquid barrier performance only
Highest tier tested bySee current textHigher limits for liquid, wet microbial and strength testsResistance to synthetic blood for drapes and to viral penetration for gowns

Different ways of classifying a barrier

The most visible difference is the classification itself.

AAMI PB70 uses four levels. Levels 1 to 3 are based on water resistance: an impact penetration test, in which water is sprayed onto the material and the amount passing through is weighed, and from Level 2 upwards a hydrostatic pressure test as well. Level 4 is a different kind of requirement. The material must resist penetration by synthetic blood in the case of drapes, and by a virus surrogate in the case of gowns, under the relevant ASTM test methods. Seams and attachment points in the critical zones are assessed, not only the base fabric.

EN 13795 uses two classes. For each class it sets limits for the critical and less critical areas across a wider group of characteristics. A gown cannot claim a class on the strength of its liquid resistance alone; it must also meet the requirements for microbial penetration, cleanliness, particle release and strength.

IS 17334:2019 likewise grades products by barrier performance, and Indian tenders commonly cite a performance level with the standard number. The definitions of those levels and the tests behind them should be read from the standard, not inferred from the other two frameworks.

What each one measures, and what it leaves out

PB70 is deliberately narrow. It classifies liquid barrier performance and says nothing about linting, strength or comfort; in the United States those are addressed by other standards. EN 13795 is broader on the product but does not test resistance to blood or to viral penetration as PB70 Level 4 does. It relies instead on a wet bacterial penetration test with mechanical rubbing.

None of the three is a sterilisation, packaging or quality management standard. A gown that meets the top tier of any framework still needs a validated sterilisation process, a sterile barrier pack and a controlled manufacturing system, each of which is governed by other standards.

Why the ratings do not convert

It is tempting to build a conversion chart, with “high performance” on one side and a level number on the other. Such charts circulate widely and should be treated with caution, for three reasons:

  • The test methods differ. A water spray test, a hydrostatic head test, a wet bacterial rubbing test and a viral penetration test challenge a fabric in different ways. A good result in one does not predict the result in another.
  • The zones differ. Each framework defines the exposed part of the product in its own terms, and seams are treated differently.
  • The scope differs. A product could reach a high liquid barrier level and still fail a requirement for linting or wet strength that the other framework imposes.

The practical rule is that a claim is valid only for the framework against which the product was actually tested. A manufacturer serving several markets tests the same product against each standard and declares each result separately.

Choosing the right reference, and what to ask your supplier

Start from the regulatory and contractual position. An Indian public tender will usually call for IS 17334. A European importer will expect EN 13795 evidence as part of the CE marking file. A buyer working to American practice will ask for an AAMI level. Where there is a choice, pick one primary framework and state the class or level for each product line.

Then ask the supplier three things: which standard and edition the product was tested to, which class or level it achieved and in which zone, and whether the report relates to the finished, sterilised product. Shrisay Medical’s Reinforced Surgical Gown is described as offering AAMI Level 3–4 protection; details of this and the rest of the surgical gown range and surgical drape range are available on request, and our quality and certifications page lists the registrations we hold.

Frequently asked questions

Is EN 13795 high performance equal to AAMI Level 4?

No direct equivalence exists. The two frameworks use different tests; AAMI Level 4 involves blood or viral penetration resistance, which EN 13795 does not test. A product must be tested to each standard separately.

Which standard should an Indian hospital specify?

IS 17334:2019 is the national standard and the usual reference in Indian public procurement. Hospitals may additionally cite EN 13795 or AAMI PB70 where their policy or accreditation requires it.

Do these standards cover sterility?

No. All three deal with barrier and material performance. Sterilisation, sterile packaging and quality management are covered by other standards and by medical device regulations.

Sourcing surgical consumables? Send us your specification or a reference sample and we will respond with product details and pricing.