Key takeaways
- ISO 11135, ISO 11137 and ISO 17665 are process standards for EO, radiation and moist heat respectively; none of them prescribes which method to use.
- Ethylene oxide is compatible with polypropylene, films and adhesives, which is why it is so widely used for non-woven drapes and gowns.
- Radiation can embrittle and discolour ordinary polypropylene over time unless a stabilised grade is specified and verified after ageing.
- Packaging is designed around the method: breathable for EO, dose-tolerant for radiation, heat- and moisture-resistant for steam.
Every sterile drape, gown or procedure pack has been through a terminal sterilisation process after it was sealed in its final packaging. Three methods account for almost all of that work in the medical-device industry: ethylene oxide gas, ionising radiation and moist heat. Each is governed by its own international standard, and each interacts differently with the polypropylene non-wovens, polyethylene films and adhesives from which single-use theatre products are made.
The method is therefore decided together with the material and the pack design, not afterwards.
Three methods, three standards
The standards share one feature that is often misunderstood: each sets requirements for developing, validating and routinely controlling a sterilisation process. None of them is a product standard, and none of them says which method a manufacturer should choose.
- ISO 11135 covers ethylene oxide (EO) sterilisation of medical devices.
- ISO 11137 covers sterilisation by radiation. It is published in parts dealing with the process requirements, the establishment of the sterilising dose and dosimetry, and it applies to gamma rays, electron beams and X-rays.
- ISO 17665 covers moist heat, the method known in every hospital as steam sterilisation or autoclaving.
The choice between them is the manufacturer’s, made on the basis of material compatibility and pack design. All three are expected to deliver the same level of sterility assurance.
How each method works
Ethylene oxide
EO is a reactive gas that inactivates micro-organisms by chemically altering their proteins and nucleic acids. The process runs at a moderately warm temperature, far below anything that would distort a plastic, with controlled humidity and a series of vacuum and gas phases. Because the gas is toxic, the load is aerated afterwards so that residues fall to acceptable levels. The process is described in more detail in our article on ethylene oxide sterilisation of single-use surgical products.
Radiation
Gamma rays from a cobalt-60 source, or electrons from an accelerator, pass through the packed product and damage microbial DNA. There is no significant heating, no gas and no chemical residue, and gamma rays in particular penetrate dense, fully sealed cartons. The dose is measured in kilogray and is established for each product from its bioburden, following the methods in ISO 11137.
Moist heat
Saturated steam under pressure transfers heat rapidly as it condenses on the load. Hospital cycles commonly run at around 121 °C or 134 °C. Steam is quick and leaves no toxic residue, which is why it remains the standard method for reusable instruments and textiles.
Why polypropylene decides the matter
Most disposable drapes and gowns are built on spunbond and meltblown polypropylene, often combined with polyethylene film, hot-melt or pressure-sensitive adhesives, elastic and absorbent layers. The way these materials respond to each method largely settles the choice.
EO leaves polypropylene essentially unchanged. The low process temperature does not soften fibres or films, repellent finishes are preserved and adhesives keep their tack. This compatibility with almost every component is the main reason EO is so widely used for non-woven surgical products.
Radiation attacks the polymer chain. In ordinary polypropylene, irradiation breaks molecular chains and starts an oxidation process that continues slowly during storage. The practical effects can be embrittlement, loss of tear and tensile strength, yellowing and a noticeable odour, and they may appear only months after the product was sterilised. Radiation-stabilised polypropylene grades are made for this purpose, but the fabric has to be specified that way from the start and its properties verified after irradiation at the maximum dose and again after ageing.
Steam is limited by heat and moisture. Polypropylene does not melt at steam temperatures, and SMS wraps designed for instrument trays are routinely autoclaved. A finished disposable is a different matter: fabrics can soften and shrink, thin polyethylene films and hot-melt adhesives may distort or flow, and tightly folded laminates trap moisture. For these reasons steam is rarely used for laminated single-use drapes and gowns.
| Method | Process standard | Effect on polypropylene non-wovens | Packaging needed |
|---|---|---|---|
| Ethylene oxide | ISO 11135 | Well tolerated by fabric, film and adhesives | Gas-permeable sterile barrier; aeration required |
| Gamma radiation | ISO 11137 | Can embrittle and discolour unless a stabilised grade is used | May be fully sealed and impermeable; materials must tolerate the dose |
| Electron beam | ISO 11137 | As for gamma; effects depend on dose and grade | As for gamma, with closer control of pack density |
| Moist heat (steam) | ISO 17665 | Softening, shrinkage and distortion of films and adhesives in finished products | Porous and resistant to heat and moisture |
Packaging follows the method
The sterile barrier system has to let the sterilising agent reach the product and then keep micro-organisms out for the rest of the shelf life. ISO 11607 sets the requirements, and the design differs by method.
- For EO, at least one face of the pouch must be gas-permeable, usually a medical-grade paper or a similar breathable web. Air has to leave, gas and humidity have to enter, and the gas has to leave again. Carton and pallet arrangements form part of the validated load.
- For radiation, the pack can be an impermeable all-film pouch, but every packaging material must itself withstand the dose, and the density of the load is mapped so that the whole pallet receives a dose between the specified minimum and maximum.
- For steam, the packaging must be porous, tolerate heat and moisture, and dry properly, since a wet pack is treated as compromised.
The method also appears on the label as a symbol defined in ISO 15223-1; see our guides to reading a sterile medical-device label and to sterile barrier packaging and shelf life.
What this means in practice
- A change of method is a design change. A product validated for EO cannot simply be sent for irradiation. Materials, packaging and the process itself would all need to be re-evaluated.
- Kits must be compatible as a whole. Every component in a procedure kit or pack, including items bought in from other makers, has to tolerate the one process applied to the sealed pack.
- Single-use sterile products are not for re-sterilisation. Putting an opened or expired drape through a hospital autoclave subjects it to a process it was never validated for. Hospital policy and the manufacturer’s instructions govern what is done with such stock.
Questions to ask a supplier
- Which sterilisation method is used, and to which standard has the process been validated?
- Is sterilisation performed in-house or by a contract steriliser, and what certification does that site hold?
- When was the process last requalified, and does the validation cover this product family and pack configuration?
- If radiation is used, is the fabric a radiation-stabilised grade, and is there data on strength after ageing?
Shrisay Medical manufactures surgical drapes, gowns and procedure packs in Faridabad to the buyer’s specification; for questions about sterilisation and packaging documentation for a particular product, please contact our team.
Frequently asked questions
Can a single-use drape be re-sterilised in a hospital autoclave?
It should not be. A single-use product is validated for one sterilisation process in its original packaging, and steam can distort films, adhesives and fabric. Follow hospital policy and the manufacturer’s instructions for opened or expired stock.
Is gamma sterilisation better than EO because it leaves no residue?
Not in general. Radiation avoids gas residues but can degrade ordinary polypropylene, while EO is gentle on materials but needs aeration. The suitable method depends on the materials and pack design.
How can I tell which method was used on a product?
The sterile symbol on the label identifies the method, using the symbols defined in ISO 15223-1. The supplier’s technical documents should state the process standard as well.