Key takeaways

  • The sterile barrier system is the minimum packaging that prevents micro-organisms from entering and allows aseptic presentation at the point of use.
  • Shelf life is established by the manufacturer through stability testing of the packaging and product, within the framework of ISO 11607.
  • A use-by date is valid only if the pack has been stored as directed and remains intact; moisture, punctures, open seals and crushing all compromise it.
  • Good storage practice in stores and theatres is the last and least controlled link in maintaining sterility.

Sterilisation makes a product sterile. Packaging keeps it that way, sometimes for years, through transport across continents, warehouse storage and handling in the hospital, until the moment a circulating nurse peels the pack open. If the packaging fails at any point, the sterilisation is undone, and usually without any visible sign on the product inside. This article explains how sterile packaging is designed, how shelf life is established, and what stores and theatre staff can do to protect it. Requirements for packaging of terminally sterilised medical devices are set out in ISO 11607; hospital storage should follow local policy and the manufacturer’s instructions.

What a sterile barrier system is

ISO 11607 uses three terms that are useful to keep apart.

  • The sterile barrier system is the minimum package that minimises the risk of ingress of micro-organisms and allows aseptic presentation of the sterile contents at the point of use. For a drape or gown this is typically a sealed pouch or a wrapped and sealed pack.
  • Protective packaging is the configuration of materials designed to prevent damage to the sterile barrier system and its contents until the point of use. Dispenser boxes and shipping cartons are examples.
  • The packaging system is the two together.

A sterile barrier system has to do several things at once. It must allow the sterilant to enter and leave, which for ethylene oxide means at least one breathable material such as medical-grade paper or a suitable non-woven. It must then keep micro-organisms out. Its seals must be continuous and strong enough to survive distribution, yet open cleanly without shedding fibres or tearing, so that contents can be delivered onto the sterile field without touching the outside of the pack.

How shelf life is established

The use-by date on a sterile pack is not an estimate. The manufacturer has to show that the packaging system maintains its integrity, and the product its performance, over the claimed period. In outline this involves:

  1. Validating the packaging process, so that sealing equipment consistently produces seals within defined limits of temperature, pressure and time.
  2. Performance testing, in which packed product is subjected to simulated distribution hazards such as vibration, compression and drops, and then examined.
  3. Stability testing, in which packs are aged and then tested for seal strength and integrity. Accelerated ageing at elevated temperature is commonly used to support an initial claim, with real-time ageing carried out to confirm it.

The shelf life also depends on the product. Adhesives on incise areas and fixation tapes, elastic components and some films change with age, so the date reflects whichever element is limiting. A longer shelf life is not inherently better; what matters is that the claim is supported by data.

What compromises a pack

HazardHow it compromises sterilityHow to prevent it
MoistureWet porous packaging can draw micro-organisms through; dried water marks show a pack has been wetStore away from sinks, pipes, external walls and floors; never place packs on wet surfaces
Punctures and tearsAny breach, however small, is a route inAvoid sharp shelf edges and overfilled bins; do not write on packs with hard-tipped pens
Open or creased sealsChannels across a seal break the barrierDo not fold or bend pouches; do not bundle with tight rubber bands
CompressionCrushing stresses seals and can force air and contaminants throughRespect stacking limits on cartons; do not stack heavy items on top
Heat and sunlightDegrade films, adhesives and elastic over timeStore within the conditions stated on the label, away from direct sun and heat sources
Dust and pestsContaminate outer surfaces and can damage packagingUse clean, closed or regularly cleaned shelving; keep shipping cartons out of clean areas
Excessive handlingEvery transfer is a chance to damage the packHandle as little as possible; rotate stock without rummaging

Storage in the hospital

General good practice for a sterile store is well established, though each hospital will have its own policy:

  • A clean, dry, well-ventilated room with controlled access, kept apart from dirty utility areas.
  • Shelving that holds packs clear of the floor, the ceiling and outside walls, with surfaces that can be cleaned.
  • Removal of outer shipping cartons before stock enters clean areas, since corrugated cartons have travelled through uncontrolled environments and shed dust.
  • Stock rotation so that the earliest use-by date is issued first.
  • Routine checks for expired or damaged stock, with a clear way to quarantine it.

Many institutions regard the sterility of an intact commercial pack as dependent on events, such as damage or wetting, in addition to the date. In that approach the use-by date is the outer limit and the condition of the pack is checked every time. Follow whichever policy your infection control team has set.

Handling from store to sterile field

The last few metres matter as much as the journey before them. Carry packs flat, not tucked under an arm. Transport them in clean, covered trolleys or containers. Inspect each pack immediately before opening: all sides, all seals, the indicator and the date. Open by peeling the seal as designed, not by pushing the contents through the paper or cutting across the pack, and present the contents in line with aseptic technique.

A pack that has been dropped on the floor, has become wet or shows any breach should be treated as non-sterile. Single-use sterile products are not to be resterilised by the hospital unless the manufacturer’s instructions expressly allow it.

What to ask your supplier

Ask what the sterile barrier system consists of, what shelf life is claimed and what stability data supports it. Ask about the storage conditions stated on the label, the stacking limits for cartons, and how many units are in each inner box so that outer cartons can be removed at the store without loose packs being handled repeatedly. For customised products, ask whether a change in fold or pack contents affects the packaging validation. Shrisay Medical supplies surgical drapes, gowns and procedure kits and packs with packaging configured to customer requirement; pack formats and labelling can be discussed through the contact page.

Frequently asked questions

Is a sterile pack still sterile if the outer carton is damaged?

The carton is protective packaging, not the sterile barrier. Inspect each unit pack inside; if the sterile barrier itself is intact, dry and undamaged, follow hospital policy, and reject any pack that shows a breach.

How is the shelf life of a sterile surgical pack decided?

The manufacturer establishes it through validation and stability testing of the packaging and product, commonly using accelerated ageing supported by real-time ageing, within the framework of ISO 11607.

What should be done with a sterile pack that has got wet?

Treat it as non-sterile and do not use it, even if it has since dried. Quarantine it and deal with it according to hospital policy.