Key takeaways
- EN 13795 is the European standard for surgical drapes, gowns and clean air suits used as medical devices, and applies to both single-use and reusable products.
- It has two performance classes, standard and high, chosen according to the duration, fluid load and mechanical stress of the procedure.
- Each product is divided into critical and less critical areas, and different tests and limits apply to each.
- A meaningful declaration states the part of the standard, the edition, the performance class and where the critical area lies.
EN 13795 is the European standard that sets requirements and test methods for surgical drapes, gowns and clean air suits used as medical devices for patients, clinical staff and equipment. It applies to single-use and reusable products alike. Anyone supplying these products to European customers, or buying from a manufacturer who quotes the standard, will meet two pairs of terms: standard performance and high performance, and critical and less critical product areas. This article explains how the two pairs fit together. It describes the framework in general terms; the limits for each test are in the standard itself and should be taken from the current edition.
How the standard is organised
The standard is published in parts. One part deals with surgical drapes and gowns. A separate part deals with clean air suits, which are garments designed to reduce the dispersal of skin particles from the wearer into theatre air and serve a different purpose from a surgical gown. A claim should therefore name the part as well as the standard number and its year.
The underlying approach is that a gown or drape is a barrier, and that a barrier is defined by measurable characteristics of the material and the finished product. The standard does not prescribe a fabric, a weight or a design. An SMS non-woven, a laminated fabric and a reusable woven textile can all be assessed against it, and each must reach the same limits to claim the same class.
Standard performance and high performance
The two performance classes reflect the fact that procedures differ in what they demand of a barrier.
- Standard performance sets the minimum requirements for products used in invasive surgical procedures. It is intended for procedures of shorter duration with limited fluid and limited mechanical stress on the material.
- High performance sets more demanding requirements, principally for liquid resistance, wet microbial penetration and strength. It is intended for procedures that are long, involve large volumes of fluid, or place heavy mechanical stress on the gown or drape.
The standard defines the classes; it does not decide which procedures need which class. That is a clinical and infection control decision for the hospital, usually made by procedure type. Arthroscopy, caesarean section and major open abdominal or orthopaedic work are commonly treated as high-demand because of fluid volume, but local policy governs.
Critical and less critical product areas
The second idea concerns the product, not the procedure. A critical product area is the part with a greater probability of being involved in the transfer of infective agents to or from the wound. On a gown this is typically the front and the sleeves. On a drape it is the area around the incision site. A less critical product area is the remainder, such as the back of a gown or the outer margins of a drape.
The manufacturer is responsible for identifying the critical area of each product. This is one reason reinforced gowns and drapes exist: an additional layer or an impervious film is placed where the challenge is highest, while the rest of the product stays lighter and more breathable.
Which tests apply to which area
The characteristics evaluated under the part covering drapes and gowns are listed below. Not every test applies to every area, which is the point most often misread.
| Characteristic | What it measures | Critical area | Less critical area |
|---|---|---|---|
| Microbial penetration, dry | Passage of micro-organisms carried on dry particles | Not required | Required |
| Microbial penetration, wet | Passage of micro-organisms through wet material under rubbing | Required | Not required |
| Microbial cleanliness | Viable micro-organisms on the product before use | Required | Required |
| Particle release | Shedding of lint and particles | Required | Required |
| Liquid penetration | Resistance to water under rising hydrostatic pressure | Required | Required, at a lower limit |
| Bursting and tensile strength, dry | Resistance to rupture and tearing | Required | Required |
| Bursting and tensile strength, wet | The same properties once the material is wet | Required | Not required |
The logic is straightforward. Critical areas are expected to get wet and to be pressed and rubbed, so they are tested wet. Less critical areas are expected to stay dry, so airborne and particle-borne transfer is the relevant route. The numerical limits differ between standard and high performance for some characteristics and are identical for others. Confirm both the applicability and the limits against the current text before relying on this summary.
Reading a declaration of performance
A useful claim under EN 13795 contains more than the number. Look for the following:
- The part and edition of the standard, since requirements and test methods have changed between editions.
- The performance class, standard or high, for the specific product code.
- A statement or drawing showing where the critical area is.
- Test evidence for both areas, from a competent laboratory, relating to the finished product as supplied, including after sterilisation where the product is sold sterile.
Remember also what the standard does not cover. It is not a sterilisation standard, a packaging standard or a quality management standard, and compliance with it does not by itself amount to CE marking. In the European Union it is one of the tools a manufacturer uses to show that a device meets the applicable regulatory requirements.
What to ask your supplier
When a quotation says “EN 13795 compliant”, ask which class, which part and edition, and which product the test report describes. Ask to see the critical area marked on the product drawing. If the product is customised, ask whether the change of fabric, reinforcement or size affects the evidence. Shrisay Medical manufactures surgical gowns and surgical drapes to customer requirement and supplies overseas buyers; requirements for a specific market can be raised through our export page.
Frequently asked questions
Does EN 13795 specify which procedures need high performance products?
No. The standard defines the two performance classes and their requirements. Choosing the class for a procedure is a decision for the hospital, based on duration, fluid volume and mechanical stress.
Who decides where the critical area of a gown or drape is?
The manufacturer identifies the critical and less critical areas for each product. Buyers should ask for this to be shown on a drawing or in the product information.
Is EN 13795 compliance the same as CE marking?
No. EN 13795 addresses product performance. CE marking in the European Union requires the manufacturer to meet the full regulatory requirements for medical devices, of which performance evidence is one part.