Key takeaways

  • Medical-device labels use internationally standardised symbols, defined in ISO 15223-1, so that key information can be understood without translation.
  • The symbols a theatre team most needs are those for sterility and method, single use, use-by date, lot number and package integrity.
  • The lot number links a pack to its production and sterilisation records and should be recorded wherever hospital policy requires traceability.
  • A pack that is damaged, wet, open or past its use-by date should not be used, whatever the rest of the label says.

Every sterile surgical pack carries a label, and most of what it says is conveyed by symbols. They are small, they are easy to overlook in a busy theatre, and they answer the questions that matter before a pack is opened: what is this, is it sterile, is it still within date, and can it be traced if something goes wrong. This article explains the common symbols and markings on single-use sterile products such as drapes, gowns and procedure packs. It is a general guide. The symbols are defined in ISO 15223-1, and national regulations add their own labelling particulars, so the manufacturer’s label and instructions always take precedence.

Why labels rely on symbols

Medical devices cross borders, and a pack made in one country may be opened by staff who read a different language. Standardised symbols allow the same label to be understood everywhere and reduce the amount of small text on a crowded pack. ISO 15223-1 sets out the symbols and their meanings. Regulators in many markets, including the European Union and India, recognise or reference them, while also requiring certain information in words.

The symbols seen most often

SymbolMeaningWhat to do with it
STERILE in a box, followed by EOSterilised using ethylene oxide. Other letters indicate other methods, such as R for irradiation.Confirms the device is supplied sterile and by which method
The figure 2 in a circle, struck throughDo not re-use; for single use onlyDiscard after one procedure according to hospital waste policy
Hourglass with a dateUse-by dateDo not use after this date
LOT in a box with a codeBatch or lot codeRecord where traceability is required; quote in any complaint
REF in a box with a codeCatalogue or reference numberIdentifies the exact product variant for reordering
Factory outline with name and addressManufacturerIdentifies who is legally responsible for the device
Factory outline with a dateDate of manufactureIndicates the age of the product
Damaged package in a circle, struck throughDo not use if the package is damagedInspect the pack before opening
Open book with the letter iConsult the instructions for useRead the accompanying or electronic instructions
Umbrella with raindrops; sun with a packageKeep dry; keep away from sunlightFollow in storage and transport

Other symbols appear depending on the product and market. These include a mark identifying the item as a medical device, a symbol accompanying the unique device identifier, a struck-through symbol meaning do not resterilise, symbols indicating whether the pack has a single or double sterile barrier, and regulatory marks such as the CE mark with the authorised representative’s details for products sold in Europe.

Lot numbers and traceability

The lot number, sometimes called the batch number, identifies a group of products made under the same conditions. For a sterile product it connects the pack in your hand to the materials used, the production record and the sterilisation cycle. It is different from the reference number: the reference tells you which product it is, and the lot tells you which production run it came from.

This distinction matters in three situations:

  • Complaints. A manufacturer can investigate a reported defect properly only if the lot number is supplied. Keep the label or the outer pack when setting a suspect product aside.
  • Recalls and field notices. These are issued by lot. Stores need to be able to find affected stock quickly.
  • Patient records. Many hospitals record the lot numbers of sterile items used in a procedure. Some packs carry peel-off labels for this purpose. Follow your own institution’s policy.

Increasingly, labels also carry a unique device identifier in a barcode or two-dimensional code. It typically combines a fixed part identifying the product with a variable part carrying the lot and expiry date, and it can be scanned into stock and patient record systems.

Dates: manufacture, use-by and shelf life

Two dates commonly appear. The date of manufacture shows when the product was made. The use-by date is the date after which the manufacturer no longer supports its use, and for a sterile product it reflects the period over which the packaging has been shown to maintain sterility and the product to remain fit for purpose under the stated storage conditions.

Dates on device labels are normally written with the year first, followed by the month and, where given, the day. A date showing only a year and month should be interpreted according to the manufacturer’s stated convention; if it is unclear, ask. The use-by date assumes that the pack has been stored as directed and is intact. A pack well within date that has been crushed, wetted or punctured is no longer reliable.

Checks before opening a pack

At the point of use, a short routine covers the essentials:

  1. Confirm the product name, size and reference are what the procedure requires.
  2. Find the sterility symbol and check the use-by date.
  3. Inspect the pack on all sides for tears, punctures, open seals, moisture or staining.
  4. Check any process indicator on the pack as described by the manufacturer.
  5. Record the lot number if required.

If any check fails, do not use the pack. Set it aside, keep the label and report it through the hospital’s system.

What to ask your supplier

When evaluating a new product, ask for the label artwork as well as the sample. Check that it carries the particulars required in your market, that the lot and use-by date are printed legibly and durably, and that the same lot number appears on the unit pack and the outer carton. For private-label orders, agree in writing whose name appears as manufacturer and who approves the artwork. Shrisay Medical offers private-label and customised packaging across its product range, including procedure kits and packs; labelling requirements for a particular market can be discussed through the contact page.

Frequently asked questions

What does STERILE EO mean on a pack?

It means the product has been sterilised using ethylene oxide. The letters after the word STERILE indicate the method, and the claim applies only while the package is intact and within its use-by date.

What is the difference between a LOT number and a REF number?

REF is the catalogue number identifying the product type and variant. LOT identifies the specific production batch, linking the pack to its manufacturing and sterilisation records.

Can a pack be used after the use-by date if it looks intact?

No. The use-by date marks the limit of the period the manufacturer has validated. Follow hospital policy for removing expired stock.

Sourcing surgical consumables? Send us your specification or a reference sample and we will respond with product details and pricing.
Request a quote
← All articles