Key takeaways
- Ethylene oxide is a low-temperature gas process suited to non-wovens, plastic films and mixed-material packs that would not tolerate steam.
- A cycle moves through preconditioning, air removal, humidification, gas exposure, gas removal and aeration, with temperature, humidity, gas concentration and time all controlled.
- Sterility cannot be confirmed by inspecting finished packs, so the process is validated and then routinely monitored, with ISO 11135 as the governing standard.
- Aeration and control of residues are as much a part of the process as the gas exposure itself.
The word STERILE on a drape or gown pack, followed by the letters EO, tells the user that the product was sterilised with ethylene oxide gas. It is the most widely used method for single-use surgical textiles and procedure packs. The process is not visible to the buyer and leaves no obvious trace on the product, so it is worth understanding how it works and how its reliability is established. This article explains the process in general terms. The development, validation and routine control of EO sterilisation for medical devices are specified in ISO 11135, and the details of any particular cycle belong to the manufacturer’s validated process.
Why ethylene oxide suits single-use surgical products
Ethylene oxide inactivates micro-organisms by reacting chemically with essential components of the cell. It does so at relatively low temperatures, which is the main reason for its use. Polypropylene non-wovens, polyethylene films, adhesives, tubing and moulded plastic components can be distorted or weakened by the heat of steam sterilisation. EO avoids that.
The gas also penetrates well. It diffuses through breathable packaging materials, into folded drapes and through the layers of a multi-component procedure pack, and then diffuses out again. This allows products to be sterilised in their final sealed packaging, already packed in shipping cartons, which removes the risk of recontamination after the process.
The other principal industrial method is irradiation. Some polymers used in surgical products discolour or lose strength when irradiated, so the choice of method is made by the manufacturer according to the materials and the pack design.
The cycle, stage by stage
Cycle designs vary, but a typical process follows this sequence:
- Preconditioning. Palletised product is held in a warm, humid room or chamber. Micro-organisms that are dry are harder to kill, and cold or dry loads respond unevenly, so the whole load is brought to a consistent temperature and moisture content.
- Air removal. The load is placed in the steriliser chamber and air is evacuated. Removing air allows gas and moisture to reach the interior of every pack.
- Conditioning. Steam is admitted to restore and maintain humidity within the load under vacuum.
- Gas injection and exposure. Ethylene oxide is introduced to the specified concentration and held for the validated exposure time at controlled temperature.
- Gas removal. The chamber is evacuated and flushed repeatedly with air or inert gas to remove ethylene oxide from the chamber and the load.
- Aeration. The load is transferred to a heated, ventilated area, or kept in the chamber, while remaining gas desorbs from the product and packaging.
Four variables govern the lethality of the process: gas concentration, temperature, humidity and exposure time. They are interdependent, and the validated cycle fixes an acceptable range for each. Because the gas is flammable and hazardous to health, the process is run in purpose-built facilities with strict engineering and environmental controls.
How sterility is assured
A sterile product cannot be verified by testing it, because testing destroys the sample and a small sample cannot prove the state of a whole load. Sterility is therefore expressed as a probability, the sterility assurance level, and is achieved by validating the process and then showing that each routine cycle stayed within the validated limits.
Validation follows the usual sequence of installation, operational and performance qualification. In performance qualification the manufacturer demonstrates, using biological indicators carrying a known population of highly resistant bacterial spores placed in the hardest-to-reach locations of the load, that the cycle delivers the required lethality with a margin of safety. Temperature and humidity sensors distributed through the load confirm that conditions are uniform.
Several features of the product itself are part of the validated state:
- The bioburden, meaning the population of micro-organisms on the product before sterilisation, which depends on raw materials and the cleanliness of manufacture.
- The product design and density, including how drapes are folded and how many components a pack contains.
- The packaging materials and the loading pattern of cartons on the pallet.
A significant change to any of these calls for an assessment of whether the validation still applies. This is relevant to customised packs: adding components or changing the fold can alter how the gas penetrates.
Aeration and residues
Ethylene oxide is absorbed by many materials and is released slowly. The sterilant itself, and compounds that can form from it, must be reduced to acceptable levels before the product is released. Limits for residues in medical devices are addressed in the ISO 10993 series on biological evaluation. Aeration time depends on the materials, the density of the load and the temperature of the aeration area, and is established during validation.
For the user this has one practical implication. Product should not reach the hospital until the manufacturer has released the lot, and release depends on completing aeration as well as the sterilisation records.
Routine release and records
Each sterilisation load produces a record of the cycle: chamber temperature, pressure, humidity, gas admission and exposure time, set against the validated specification. Depending on the release method, biological indicators from the load are also incubated and read. Chemical indicators on cartons or packs change colour on exposure to the process; they show that a pack has been through a cycle, not that it is sterile. The sterilisation lot is linked to the product lot number on the label, which is what makes a pack traceable to its cycle.
What to ask your supplier
A buyer does not need the full validation file, but a few questions are reasonable. Is sterilisation carried out in-house or by a contract steriliser, and is the process validated to ISO 11135? Does the validation cover the specific pack configuration being ordered? How is a lot released, and can a sterilisation certificate be supplied with each lot? What shelf life has been established for the sterile pack? Shrisay Medical supplies drapes, gowns and customised procedure kits and packs; sterility requirements can be confirmed for each product at the enquiry stage, and our quality and certifications page gives further background.
Frequently asked questions
Why is ethylene oxide used instead of steam for disposable drapes and gowns?
EO works at low temperature, so it does not damage non-woven fabrics, plastic films and adhesives. It also penetrates sealed breathable packaging, allowing products to be sterilised in their final packs.
Does a colour-change indicator on the pack prove sterility?
No. A chemical indicator shows that the pack has been exposed to the process. Sterility assurance comes from the validated cycle and the records for that sterilisation lot.
Can EO-sterilised single-use products be resterilised by the hospital?
Single-use products should not be reprocessed or resterilised unless the manufacturer’s instructions expressly permit it. Follow the label, the instructions for use and hospital policy.