Key takeaways
- AQL is the worst process average considered tolerable over a series of lots, not a permitted percentage of defective product.
- Lot size and inspection level give a code letter, which together with the AQL gives the sample size and the accept and reject numbers.
- Defects are classed as critical, major or minor, with a stricter limit for the more serious classes.
- Sterility and barrier performance are not controlled by AQL sampling; they rely on process validation and laboratory testing.
Specifications for drapes and gowns often contain a line such as “AQL 2.5 for major defects”, and inspection reports quote sample sizes and acceptance numbers that seem to come from nowhere. They come from acceptance sampling, a statistical method for deciding whether to accept a lot without examining every unit in it. The method is simple to apply and widely misunderstood, particularly in what an AQL figure does and does not promise.
Why sample at all
Inspecting every unit is costly, slow and, for sterile products, impossible once the pack is sealed: a drape cannot be unfolded and checked without destroying what is being sold. Acceptance sampling replaces it with a defined procedure. A random sample of stated size is drawn from the lot, each unit is examined, the defects are counted, and the lot is accepted or rejected by comparing the count with fixed numbers.
The standard most commonly used is ISO 2859-1, which covers sampling schemes for inspection by attributes, indexed by acceptance quality limit. The corresponding Indian Standard is IS 2500 (Part 1). “By attributes” means that each unit is judged as conforming or not, as opposed to measuring a quantity and analysing the values.
What AQL means
AQL stands for acceptance quality limit. It is defined as the worst process average that can be considered tolerable when a continuing series of lots is submitted.
The sampling plans are constructed so that lots from a process running at or better than the AQL will be accepted most of the time. Three consequences follow.
- It is not a permitted defect rate. An AQL of 2.5 does not entitle a supplier to ship 2.5 per cent defective product. The supplier’s process is expected to run better than the AQL.
- It describes a series of lots. The scheme is designed for continuing production, and its protection comes from applying it lot after lot.
- It does not guarantee any individual lot. A sample can pass by chance when the lot is worse than the AQL, and the smaller the sample, the larger that risk.
From lot size to a decision
Using the tables in the standard follows a fixed sequence.
| Step | What is decided | Where it comes from |
|---|---|---|
| 1. Lot size | The number of units in the lot offered for inspection | The supplier’s lot definition |
| 2. Inspection level | How large the sample is relative to the lot | Specification or contract; general level II is the usual default |
| 3. Code letter | A letter corresponding to lot size and inspection level | The code-letter table in the standard |
| 4. AQL | The quality limit for each class of defect | Specification or contract |
| 5. Sample size and acceptance numbers | How many units to inspect, and the accept and reject numbers | The master table for the code letter and AQL |
| 6. Severity | Normal, tightened or reduced inspection | Switching rules based on recent lot history |
The standard provides three general inspection levels, I, II and III, in increasing order of sample size, and four special levels, S-1 to S-4, which give small samples for use where inspection is destructive or expensive.
As an illustration, for a lot of between 1,201 and 3,200 units at general level II, the tables give code letter K and a sample of 125 units under normal single sampling. At an AQL of 1.0, the lot is accepted if three or fewer nonconforming units are found and rejected if four or more are found. The figures are quoted only to show how the tables work; a plan for actual use should be taken from the current edition of the standard.
Sample size does not rise in proportion to lot size, so a large lot is sampled at a lower percentage than a small one. The switching rules also matter: after a run of rejections the scheme moves to tightened inspection, and returns to normal only after a sustained run of acceptances.
Defect classes, with examples
Not all defects matter equally, so they are classified and each class is given its own AQL: the more serious the class, the stricter the limit. The classification belongs in the specification; the examples below are illustrative, not a universal list.
Critical
A defect likely to cause harm or to make the product unsafe to use.
- An open or incomplete seal on the sterile pack, or a hole in the pouch.
- A hole or tear in the barrier area of a drape, or in the critical zone of a gown.
- A wrong or missing sterile label, lot number or expiry date.
- Foreign matter, such as an insect or a metal fragment, inside the pack.
Major
A defect likely to cause failure in use or to reduce usability materially.
- A fenestration of the wrong size or in the wrong position.
- An adhesive strip that is missing or has no release liner.
- An open sleeve seam, a missing tie or a detached cuff on a gown.
- A wrong component, or a component missing, in a kit.
Minor
A departure from specification with little effect on use.
- A slight variation in shade.
- Untidy thread ends or a marginally uneven fold.
Many specifications allow no critical defects in the sample at all, treating any such finding as grounds for rejection and investigation, though a zero count in a sample cannot prove that the lot contains none.
Why AQL does not apply to sterility
Sterility cannot be handled as an inspection attribute with an AQL. The assurance required for a sterile product is a probability of the order of one in a million, far beyond anything a sampling plan can detect, and there is no acceptable percentage of non-sterile packs. Sterility is assured by a validated sterilisation process and by monitoring every load, as described in our article on ethylene oxide sterilisation.
The same caution applies to performance properties. Barrier resistance and strength are established by laboratory tests under product standards such as IS 17334:2019, which set their own sampling and acceptance criteria. AQL inspection is suited to what can be seen, counted or measured on the unit: workmanship, dimensions, assembly, labelling and pack integrity.
Using AQL in a specification
- Name the standard, the inspection level and whether single or double sampling is used.
- List the defects in each class and give the AQL for each class.
- State where inspection takes place: at the factory before sterilisation, on finished goods, or on receipt.
- State what happens to a rejected lot, such as full sorting, rework or return, and how it is resubmitted.
For sterile goods, sample inspection on receipt consumes packs, so many buyers rely on the manufacturer’s inspection records and supplier audits. Related drafting points are covered in our guides to writing a tender specification and evaluating a manufacturer.
Shrisay Medical manufactures surgical drapes, gowns and procedure packs in Faridabad to the buyer’s specification; to discuss inspection requirements for an order, please contact our team.
Frequently asked questions
Does AQL 2.5 mean that 2.5 per cent of the units may be defective?
No. It is the quality limit around which the sampling plan is designed, and the supplier’s process is expected to run better than it. It is not an entitlement to ship defective units.
Which inspection level should be specified?
General inspection level II is the usual default in ISO 2859-1. Other levels may be agreed where more or less discrimination is needed, and special levels are used when inspection is destructive.
Can sterility be accepted on an AQL basis?
No. No sampling plan can demonstrate the assurance level required for a sterile product. Sterility depends on a validated process and the records of each sterilisation load.