Key takeaways

  • ISO 10993-7 sets allowable limits for ethylene oxide and ethylene chlorohydrin residues, the methods for measuring them and the routes to product release.
  • Allowable residues depend on the device’s contact category and duration, so there is no single figure that applies to every product.
  • Aeration is a validated stage of the process, and its duration depends on materials, packaging, load density, temperature and air exchange.
  • Residual data apply to a defined product, pack and cycle, and a kit is governed by its slowest-releasing component.

Ethylene oxide is the sterilising agent best suited to polypropylene non-wovens, films and adhesives, which is why so many single-use drapes, gowns and packs are processed with it. The property that makes it effective, its readiness to react with biological molecules, also makes it hazardous to people. Some of the gas remains in the product and its packaging when the cycle ends, and the manufacturer has to show that what is left has fallen to an acceptable level before the goods are released. ISO 10993-7 is the standard that defines how.

Where the standard sits

ISO 10993 is the series on biological evaluation of medical devices, and Part 7 deals specifically with ethylene oxide sterilisation residuals. It is the companion to ISO 11135, which governs the sterilisation process itself: ISO 11135 is concerned with whether the product is sterile, and ISO 10993-7 with whether the sterilised product is safe with respect to what the gas leaves behind. The process is described in our article on ethylene oxide sterilisation of single-use surgical products.

In broad terms, Part 7 does three things. It sets allowable limits for residues, it describes how residues are to be measured, and it sets out how a manufacturer may demonstrate compliance in order to release product.

The residues in question

  • Ethylene oxide (EO) — the unreacted gas absorbed into fibres, films, adhesives and packaging during exposure, which then diffuses out over time.
  • Ethylene chlorohydrin (ECH) — a reaction product formed when ethylene oxide meets chloride ions in a material. Whether it forms, and how much, depends on the composition of the device and its packaging. It is less volatile than EO, so aeration removes it less readily.
  • Ethylene glycol (EG) — formed when ethylene oxide reacts with water. The standard discusses it, but its limits are set for EO and ECH.

Because ECH does not simply air off in the way EO does, it is controlled mainly by the choice of materials and by process conditions, not by longer aeration alone.

How the limits are framed

The standard does not set one figure for all devices. The allowable residue depends on how the device contacts the body and for how long, using duration categories that correspond to those of ISO 10993-1. Limits are expressed as the amount of residue a patient may receive from the device over defined periods, and they become more restrictive in total as contact becomes longer.

For devices that contact the body surface, which includes drapes and gowns, the standard also addresses the potential for local irritation, with a separate criterion related to the contact area. It contains provisions for special situations as well, and it has been amended since first publication, among other things to address particularly vulnerable patients such as neonates and infants.

Numerical limits are deliberately not reproduced here. They differ by category and have been revised, and anyone writing a specification or reviewing a report should work from the current edition and its amendments. What a buyer needs to understand is the logic: a short-contact surface device and a long-term implant are not held to the same figure, and a report should state which category was applied.

Why aeration follows the cycle

At the end of exposure, the chamber is evacuated and flushed, which removes the gas from the free space but not what has dissolved into the materials. That fraction leaves by diffusion, and diffusion takes time. Aeration is the controlled stage that allows for it: the load is held in a ventilated chamber or room, usually at an elevated temperature and with frequent air changes, for a period established during validation.

Aeration is part of the validated process, not a waiting period that can be shortened when an order is urgent. It also protects warehouse and hospital staff, since a load released too early would continue to give off gas inside closed cartons and storerooms.

What affects how fast residues dissipate

FactorEffect on residuesWhat the manufacturer controls
MaterialPolymers differ widely in how much gas they absorb and how quickly they release it; some plastics, rubbers and cellulose-based materials retain more than polyolefin non-wovensSelection of materials and of bought-in kit components
PackagingGas must escape through the breathable part of the sterile barrier, then through the cartonPouch design, area of permeable web, carton and overwrap
Load densityTightly folded product and closely stacked pallets slow the exchange of airFold pattern, carton fill, pallet configuration and spacing
TemperatureWarmer conditions speed diffusion; cold storage slows itHeated aeration and control of seasonal variation
Air exchangeFresh air carries away released gas and keeps the process movingVentilation rate and airflow through the load
Cycle parametersGas concentration, exposure time and post-exposure flushing set the starting levelCycle design within the validated limits

The practical consequence is that residue behaviour is specific to a product and its pack. A plain SMS gown in a paper-film pouch and a custom procedure pack containing tubing, bowls, gauze and laminated drapes in a dense wrap will not behave alike, and the pack is governed by its slowest component.

How compliance is demonstrated

Residues are measured by extracting them from the product and analysing the extract, most commonly by gas chromatography. The standard describes two approaches to extraction: one that simulates how the device is used, and exhaustive extraction, which recovers the total residue present. The choice depends on the contact category.

For release, the standard allows more than one route. A manufacturer may build dissipation curves, measuring residues at intervals after sterilisation to establish the aeration time after which product reliably meets the limits, and then release routine loads once that time and the specified conditions have been met. Alternatively, residues may be tested on individual batches. Either way the data relate to a defined product family, pack configuration and cycle, so significant changes to any of these call for reassessment.

What buyers can reasonably ask

  1. Has the product family been evaluated to ISO 10993-7, and which edition and contact category were used?
  2. Do the residual data cover both EO and ECH?
  3. Is release based on a validated aeration time or on batch testing?
  4. For kits, does the evaluation cover the complete pack, including components bought in?
  5. Is aeration carried out by the manufacturer or by a contract steriliser, and how is it recorded for each load?
  6. How are changes to materials, packaging or load pattern assessed for their effect on residues?

It is also sensible to respect the lead time that aeration requires when planning orders, and to store cartons in a ventilated area in line with the guidance in our article on sterile barrier packaging and storage.

Shrisay Medical manufactures surgical drapes, gowns and procedure packs in Faridabad for hospitals, distributors and export buyers; for documentation relating to a specific product, please contact our team.

Frequently asked questions

Can aeration be shortened to meet an urgent order?

It should not be. Aeration time is established during validation to bring residues within the limits, and releasing a load early removes that assurance.

Why does the article not give the residue limits?

The limits vary with the contact category and duration of the device and have been amended over time. They should be taken from the current edition of ISO 10993-7.

What is ethylene chlorohydrin and why does it matter?

It is a compound formed when ethylene oxide reacts with chloride ions in a material. It is less volatile than ethylene oxide, so it is controlled mainly through material choice and process design.