Key takeaways

  • ISO 10993-1 requires a risk-based biological evaluation of the finished device, not a fixed list of tests.
  • Drapes and gowns are surface-contacting devices of limited duration, in contact with intact skin and, near the incision, with breached surfaces.
  • Cytotoxicity, sensitisation and irritation are the endpoints usually at the centre of the evaluation for these products.
  • Adhesives, elastic, inks and sterilisation residues deserve closer scrutiny than the base polypropylene fabric.

A surgical drape lies on a patient’s skin for the length of an operation, with its adhesive edge fixed beside the incision. A gown is worn against the neck, wrists and forearms of the team for the same period, often over damp skin. Neither product is implanted or enters the bloodstream, yet both are in sustained contact with the body, and a material that irritates or sensitises would be a real problem in theatre. Biocompatibility evaluation is the discipline that addresses this, and ISO 10993 is the family of standards that structures it.

What ISO 10993 is, and what it is not

ISO 10993 is a series titled “Biological evaluation of medical devices”. Part 1 sets the framework: biological evaluation is to be carried out within a risk management process, of the kind described in ISO 14971. The remaining parts cover individual subjects, including tests for cytotoxicity, sensitisation and irritation, sample preparation, chemical characterisation of materials and ethylene oxide residuals.

Three points are worth stating plainly.

  • It is not a checklist of mandatory tests. Part 1 asks the manufacturer to identify the biological risks relevant to the device and to address each one, using existing data where it is adequate and testing where it is not.
  • It applies to the finished device. The evaluation concerns the product as supplied, after converting, printing, bonding and sterilisation, not just the raw fabric.
  • There is no such thing as an “ISO 10993 certificate”. The outcome is a documented evaluation, supported by test reports for particular endpoints. A statement that a product is “ISO 10993 certified” should prompt a request for the underlying reports.

The standard has been revised several times and the detail of its categories has changed between editions, so the current edition should always be consulted for specification purposes.

How contact is categorised

Part 1 classifies a device by the nature of its contact with the body and by how long that contact lasts. For drapes and gowns the relevant category is the surface-contacting device, and within it two situations arise:

  • Intact skin — the greater part of a drape resting on the patient, a disposable sheet, and the inner face of a gown against the wearer.
  • Breached or compromised surfaces — the edge of the fenestration and any adhesive or film lying at, or immediately beside, the incision.

Duration is assessed separately. A drape or gown is used for one procedure and then discarded, which places it in the shortest duration band, described in the standard as limited contact. The manufacturer must still consider foreseeable use, such as long operations and the same staff member wearing gowns day after day.

The endpoints usually considered

For surface contact of limited duration, three biological endpoints are generally at the centre of the evaluation.

  • Cytotoxicity (ISO 10993-5). An in vitro screen in which cultured cells are exposed to the material or an extract of it. It is sensitive and shows whether anything leaching from the product is harmful to cells.
  • Sensitisation (ISO 10993-10). An assessment of whether the material can provoke an allergic response on repeated exposure. This is the endpoint most relevant to adhesives, rubber components and certain chemical additives.
  • Irritation (ISO 10993-23). An assessment of whether the material causes a local inflammatory reaction on contact. Irritation was formerly dealt with alongside sensitisation and now has its own part, which also provides for in vitro methods.

For contact with breached surfaces, the framework points to further considerations, and depending on the edition and the risk assessment these can include systemic effects and pyrogenicity. Chemical characterisation of the materials, under ISO 10993-18, increasingly forms the first step, because knowing what a material contains allows many questions to be answered without biological testing.

Where the risks actually sit

Medical-grade polypropylene and polyethylene have a long record of safe skin contact. The components that deserve closer attention are the ones added to them.

ComponentTypical contactPoints to consider
Base non-woven and filmIntact skin of patient or wearerPolymer grade, repellent or antistatic finishes, residual processing aids
Adhesive strips and fenestration bordersSkin at and beside the incisionSensitisation and irritation; skin trauma on removal; residue left on the skin
Incise filmPrepared skin, cut through at the incisionAdhesive composition; any antimicrobial agent incorporated in the film
Cuffs and elasticWearer’s wrists, under glovesWhether natural rubber latex is present, and how this is labelled
Inks and colourantsUsually indirectPrinted marks kept off skin-contact faces; pigments fixed in the polymer
Sterilisation residuesAll contact surfacesEthylene oxide residuals, addressed under ISO 10993-7

Adhesives are the most frequent source of complaints. An adhesive has to hold on skin prepared with antiseptic and still release cleanly, and skin reactions to adhesive products are a recognised clinical issue. The different roles of adhesive edges and full adhesive films are covered in our article on incise drapes and adhesive films.

Why a change of material matters

A biological evaluation is valid for the product as it was evaluated. A new adhesive supplier, a different pigment, an added finish or a change of sterilisation method can each alter what reaches the skin. A manufacturer working under a quality system such as ISO 13485 is expected to assess such changes and record whether the existing evaluation still applies or further work is needed. For buyers of private-label or custom products, this is a reason to agree in advance how material changes will be notified.

What to ask a supplier

  1. Is there a documented biological evaluation for this product family, and which contact category and duration does it assume?
  2. Which endpoints were tested, to which parts and editions of ISO 10993, and by which laboratory?
  3. Were the tests carried out on the finished, sterilised product, or on representative materials?
  4. What adhesive is used, and is there separate data for it?
  5. Does the product, including cuffs and elastic, contain natural rubber latex?
  6. How are changes of raw material controlled and communicated?

Biocompatibility data does not replace clinical judgement. Patients and staff with known sensitivities to adhesives or other materials are managed under hospital policy, and any suspected reaction to a product should be reported to the supplier with the lot number. Our guide to evaluating a surgical drape manufacturer places these questions in the wider context of supplier approval.

Shrisay Medical manufactures surgical drapes and gowns in Faridabad to order; for information on the materials and adhesives used in a particular product, please contact our team.

Frequently asked questions

Is there an ISO 10993 certificate for a surgical drape?

No. ISO 10993 leads to a documented biological evaluation supported by test reports for specific endpoints. Ask for those reports and check which product and edition they refer to.

Are biocompatibility tests done on the fabric or on the finished product?

The evaluation should address the finished product as supplied, including adhesives, printing and the effects of sterilisation. Tests on raw fabric alone may not cover everything that contacts the skin.

Does a biocompatibility report guarantee that no patient will react?

No. It shows that the materials meet accepted criteria, but individuals can still be sensitive to adhesives or other components. Known sensitivities are managed under hospital policy.