Key takeaways
- A lot number is a key to records that link raw materials, production, packing, sterilisation load, release and dispatch.
- ISO 13485 requires documented traceability procedures and batch records, including sterilisation parameters traceable to each production batch.
- Traceability is often lost at the distributor, so lot numbers should be recorded on every onward supply.
- In a recall, the hospital’s task is to identify, quarantine and account for the stated lots and to confirm the action taken.
The short string of characters printed beside the LOT symbol on a drape pouch is easy to ignore, and on most days nobody needs it. Its value appears when something goes wrong: a seal found open in several packs, a complaint of a skin reaction, a sterilisation record that fails review. At that point the lot number is the only thing that allows a manufacturer to say which packs are affected and which are not, and allows a hospital to find them on its shelves. Without it, the choice lies between doing nothing and withdrawing everything.
What a lot number represents
A lot, or batch, is a defined quantity of product made under essentially the same conditions, so that it can be treated as uniform. For a sterile disposable this normally means the same product code, the same materials and production period, and exposure to the same sterilisation process. How large a lot is, and exactly where one ends and the next begins, is the manufacturer’s decision, set out in its procedures.
The number itself is only a key. Its usefulness depends entirely on the records filed under it. Lot numbers, expiry dates and the other label symbols defined in ISO 15223-1 are explained in our guide to reading a sterile medical-device label.
The chain a lot record should hold
Traceability runs in two directions. Backward traceability goes from a finished pack to what went into it. Forward traceability goes from a lot to the customers who received it. A complete record supports both.
| Stage | Typical record | What it links |
|---|---|---|
| Incoming material | Supplier lot or roll numbers, inspection results, supplier certificates | Fabric, film, adhesive tape and packaging to their source |
| Production | Batch manufacturing record: date, line, operators, quantities, in-process checks | Material lots to the finished lot number |
| Packing and labelling | Sealing parameters, label reconciliation, pack inspection | Product to its label, lot and expiry |
| Sterilisation | Load or cycle number, cycle records, indicator results, aeration record | Finished lot to the sterilisation load it travelled in |
| Release | Review and signature by an authorised person | All of the above to a release decision |
| Dispatch | Invoice or delivery note with lot numbers and quantities | Lot to the consignee |
The relationship between production lots and sterilisation loads needs particular care. One load may contain several lots, and one large lot may be split across loads. The records must make it possible to answer both questions: which lots were in this load, and which loads did this lot pass through.
For kits, the record should also show the lot of each bought-in component.
What ISO 13485 expects
ISO 13485, the quality management standard for medical-device manufacturers, treats identification and traceability as requirements in their own right. In outline it expects the manufacturer to:
- identify product by suitable means throughout production, and identify its status with respect to inspection and test, so that only product that has passed is released;
- document procedures for traceability, defining its extent and the records to be kept, in line with applicable regulatory requirements;
- maintain a record for each batch that shows the quantity manufactured and the quantity approved for distribution, and that is verified and approved;
- record the sterilisation process parameters for each sterilisation batch, traceable to the production batches concerned;
- retain records for at least the lifetime of the device as the manufacturer defines it, subject to the minimum period set in the standard and any longer period required by regulation.
More stringent traceability requirements apply to implantable devices than to products such as drapes and gowns. What certification does and does not tell a purchaser is discussed in our article on ISO 13485 for medical-device buyers.
The link in the middle: distributors
Traceability is often lost between the factory gate and the hospital store. A manufacturer knows which lots went to a distributor; unless the distributor records lot numbers on its own outward invoices or delivery notes, nobody knows which hospital received them. Distributors and stockists should therefore:
- record lot numbers and quantities for goods received and goods supplied;
- keep products in their original labelled cartons and avoid mixing lots in one carton;
- keep contact details of customers current, so that a notice can be passed on without delay.
What a hospital should record
Practice varies between institutions and is set by hospital policy. The following records are commonly kept and make any later investigation far simpler.
- At receipt — product, lot number, expiry date, quantity and supplier, entered in the stores system.
- At issue — which lots went to which theatre or department.
- At the point of use — for procedure packs and other items the hospital chooses to track, the lot number noted in the theatre record. Many packs carry peel-off lot labels for this purpose.
- When a problem is found — the lot number, a description or photograph, and the affected pack and its outer packaging retained where safe to do so.
A complaint that reaches the manufacturer with a lot number can be investigated against the batch record and retained samples. One without it usually cannot be taken beyond a general review. Handling of pack defects at the point of use is covered in our article on sterile barrier packaging, storage and pack damage.
How a recall or field safety notice works
A recall is the removal or correction of a product already distributed, undertaken because it may be defective or unsafe. In India the Medical Devices Rules 2017 place obligations on manufacturers and importers in relation to recall and to informing the licensing authority, and adverse events can be reported through the national materiovigilance programme. Other markets have equivalent systems, and the term field safety notice is widely used for the communication sent to users. The usual sequence is:
- A signal arises from a complaint, an internal finding, a supplier notification or a regulator.
- The manufacturer investigates, assesses the risk and identifies the affected lots from its records.
- A notice is issued to consignees stating the product, the lot numbers, the nature of the problem and the action required.
- Distributors pass the notice to their own customers.
- The hospital checks its stock, quarantines affected lots, and returns or disposes of them as instructed.
- Recipients acknowledge the notice and report quantities, so that the manufacturer can reconcile what was recovered against what was shipped.
- The cause is addressed through corrective action, and the authority is informed as the rules require.
A notice is not always a withdrawal. It may ask users to inspect packs before use or correct a labelling error. Whatever the instruction, a hospital should act on the lot numbers stated and confirm in writing.
Shrisay Medical manufactures surgical drapes, gowns and procedure packs in Faridabad for hospitals, distributors and export buyers; for any query relating to a specific lot, please contact our team with the lot number from the pack.
Frequently asked questions
Is a lot number the same as a sterilisation load number?
Not necessarily. A sterilisation load may contain several production lots, and a large lot may be split across loads. The manufacturer’s records should link the two in both directions.
Does a hospital have to record the lot number of every drape used?
That is decided by hospital policy. Recording lots at receipt and issue is common, and noting the lot of procedure packs in the theatre record makes later investigation much easier.
What should a hospital do on receiving a field safety notice?
Check stock against the lot numbers stated, quarantine any affected packs, follow the instructions in the notice and return the acknowledgement with the quantities found.