Key takeaways
- A drape is folded so that it can be placed on the site first and then opened outwards without the user touching the patient side.
- Fan-folds pay out in one continuous movement, book-folds open leaf by leaf, and large drapes usually combine the two.
- Arrows, head and foot markers and a cuff at the edge help the team orient the drape and protect gloved hands.
- Pack contents are stacked in order of use, and the stacking order should be specified along with the contents.
Unpacked and laid flat, a laparotomy drape may be longer than the operating table and wider than two people can comfortably span. In use, it has to go from a bundle the size of a folded newspaper to its full extent in a few seconds, handled by one or two scrubbed people who may touch only one side of it, and it must land in the right place the first time. None of that is possible unless the drape has been folded for it.
What the fold has to achieve
A drape separates a sterile upper surface from an unsterile patient side. A good fold sequence lets the team apply it while respecting that separation. In practical terms the fold should make it possible to:
- identify the centre or fenestration and place it on the operative site before anything else is opened;
- open the drape outwards, from the site towards the periphery, so that fabric which has been near the unsterile edges is never drawn back across the field;
- keep gloved hands on the sterile outer surface throughout, with no need to reach underneath;
- keep the bulk of the drape gathered and under control until it is released, so that it does not trail below table level or brush against lights and equipment.
Once a drape has been placed it is not shifted towards the incision, a point covered in our refresher on the principles of the sterile field, so first-time placement has to be easy.
Fan-fold and book-fold
Most drape folding is built from two basic folds.
In a fan-fold, also called an accordion or concertina fold, the fabric is pleated back and forth on itself in equal panels. When the free edge is pulled, the pleats pay out one after another in a single smooth movement.
In a book-fold, the fabric is folded in halves, like closing a book, and often in halves again. It opens leaf by leaf. Each opening is a deliberate action, which gives control over when each section is released but requires more handling.
Large drapes normally use both. A common arrangement is to fan-fold the drape along its length from each end towards the fenestration, and then to fold the resulting strip in from the sides. In theatre the sequence runs in reverse: the bundle is placed on the site, opened to the sides, and then drawn out towards the head and the foot.
| Fold type | How it opens | Typical use |
|---|---|---|
| Fan-fold | Pulled out continuously from one edge; pleats release in turn | Length of large procedure drapes; table covers |
| Book-fold | Opened leaf by leaf | Width of large drapes; plain sheets and utility drapes |
| Combined fan and book | Opened to the sides, then drawn out to head and foot | Fenestrated procedure drapes |
| Rolled or telescoped | Unrolled or extended along a limb or cable | Stockinette, leggings, camera and equipment covers |
Opening away from the site
The governing idea is that a drape travels in one direction only: from the clean centre to the edges. Folding puts the fenestration, or the centre line of a plain drape, on the outside of the bundle where it can be seen and positioned.
The patient side matters equally. The folds are arranged so that the face which will lie against the patient is turned inwards or downwards within the bundle. As the drape pays out, that face unrolls directly onto the patient and the table, and the scrubbed person’s hands remain on the upper surface.
Shaped drapes follow the same logic. U-drapes and split sheets, for example, are folded so that the tails can be passed around a limb, as described in our article on orthopaedic draping systems.
Direction arrows and head or foot markers
A folded drape gives few visual clues about which way round it goes. Manufacturers therefore add markers on the outer face of the bundle, such as:
- arrows showing the direction in which to unfold, sometimes numbered in sequence;
- the words “head” and “foot”, or a simple outline of a patient, to show orientation;
- a mark at the centre of the fenestration or the centre line;
- labels or coloured tabs on liner ends and on parts intended to be grasped.
A simple trial is to ask someone unfamiliar with the drape to apply it using only the markings.
The cuff that protects gloved hands
The outer edges of a drape end up at the boundary between the sterile field and its unsterile surroundings, and a hand carrying the edge there is at risk of touching something unsterile.
The answer is a cuff. The edge of the drape is folded back on itself so that the scrubbed person can slide gloved hands inside the fold. The fabric then covers the fingers and the backs of the hands as the drape is carried outwards, and whatever is touched at the periphery is touched by the drape, not the glove. Some drapes have a cuff built into the fold; with plain sheets the user forms one by turning the edge over the hands.
Order of items inside a pack
The same thinking applies one level up, to the way a procedure pack is assembled. Items are stacked so that the first thing needed is on top and the last is at the bottom, and nobody has to dig through sterile contents.
A typical sequence, which varies with the pack and the hospital’s practice, runs as follows.
- The outer wrap, which is often the back table cover, is opened by the circulating nurse to create the sterile surface.
- Hand towels and gowns lie on top, since the team dries and gowns before touching anything else.
- The Mayo stand cover and any equipment covers come next.
- Utility drapes or towels for squaring off the site follow.
- The main procedure drape lies beneath them, with accessories such as pouches and tube holders.
When a hospital specifies a custom pack, the stacking order deserves as much attention as the contents list. The broader case for packs is set out in our article on custom procedure packs.
Folding as a controlled manufacturing step
For a manufacturer, the fold sequence is part of the product specification. It is documented step by step, usually with diagrams, and operators are trained against it, because much folding of large drapes is done by hand on folding tables.
The fold also has to suit the packaging and the sterilisation process. The bundle must fit its pouch without strain on the seals, and it should not be compressed so tightly that sharp creases damage films or adhesive strips. Any change to the fold, however minor it appears, should therefore go through the manufacturer’s change control and be communicated to users.
Shrisay Medical makes drapes to order in Faridabad, and fold sequence and pack order can be agreed with the buyer along with size and configuration; the range is shown on our surgical drapes page.
Frequently asked questions
What is the difference between a fan-fold and a book-fold?
A fan-fold is pleated back and forth so that it pulls out in one movement. A book-fold is folded in halves and opens leaf by leaf, giving more control at each step.
Why do drapes have a cuff at the edge?
The cuff lets the scrubbed person tuck gloved hands inside the fold, so the drape and not the glove meets any unsterile surface at the edge of the field.
Can a hospital ask for a different fold or pack order?
In made-to-order products this can usually be discussed. Any change should be trialled by the theatre team and documented in the product specification.