Key takeaways

  • Under the EU Medical Device Regulation, surgical drapes and gowns are medical devices, and those supplied sterile need a notified body to assess the aspects relating to sterility.
  • The CE mark is the visible end of a process that includes technical documentation, a declaration of conformity and, for non-EU manufacturers, an authorised representative.
  • Importers carry their own legal obligations, including verification of the device, identification on the product, registration and complaint handling.
  • Transitional arrangements and guidance have changed over time, so every check should be made against the current official position.

In the European Union, surgical drapes and surgical gowns are regulated as medical devices under Regulation (EU) 2017/745, the Medical Device Regulation or MDR. The CE mark on a pack is the manufacturer’s statement that the device meets the regulation. For an importer bringing products in from a manufacturer outside the Union, that mark is where the checking begins, because the regulation gives importers duties of their own. This article outlines the framework and the points worth verifying. It is general information, not legal advice; the regulation, its guidance and its transitional provisions should be checked in their current form with a regulatory adviser.

How drapes and gowns are classified

The MDR sorts devices into risk classes using a set of classification rules. Non-invasive devices such as surgical drapes and gowns generally fall into the lowest class, Class I. The route to the CE mark then depends on whether the product is sterile.

  • Class I, non-sterile. The manufacturer carries out the conformity assessment under its own responsibility, compiles the technical documentation and issues a declaration of conformity. No notified body is involved.
  • Class I, supplied sterile. A notified body must be involved, but its assessment is limited to the aspects concerned with establishing, securing and maintaining sterile conditions. The CE mark is then accompanied by the notified body’s identification number.

Classification is the manufacturer’s responsibility and depends on the intended purpose stated for the device. One further point applies to gowns. Where a gown is intended to protect the wearer as well as the patient, personal protective equipment legislation may also be relevant. Ask the manufacturer how the intended purpose has been defined.

What stands behind the CE mark

The mark is supported by a body of documentation that the manufacturer must hold and keep up to date. In outline this includes a description of the device and its intended purpose, design and manufacturing information, the general safety and performance requirements and how each is met, risk management, verification and validation data, clinical evaluation, and a plan for post-market surveillance. For drapes and gowns, performance is normally demonstrated using EN 13795, sterilisation using the standard for the relevant process, and packaging using the standards for sterile barrier systems.

A manufacturer established outside the Union must also appoint an authorised representative within it. The representative’s name and address appear on the label, and it holds a copy of the documentation for the authorities.

The importer’s own obligations

The MDR defines several economic operators: manufacturer, authorised representative, importer and distributor. An importer is the person established in the Union who places a device from a third country on the Union market. In general terms the importer is expected to:

  1. Verify, before placing a device on the market, that it is CE marked, that a declaration of conformity has been drawn up, that an authorised representative has been designated, that the labelling and any instructions comply, and that a unique device identifier has been assigned where applicable.
  2. Indicate its own name and address on the device, its packaging or an accompanying document, without obscuring the manufacturer’s label.
  3. Complete the registration steps that apply to importers and check that the device has been registered.
  4. Ensure that storage and transport conditions under its control do not compromise the device.
  5. Keep a register of complaints, nonconforming devices, recalls and withdrawals, and pass relevant information to the manufacturer and its representative.
  6. Refrain from supplying a device it believes is not in conformity, inform the manufacturer, and co-operate with the competent authorities.

An importer who markets a device under its own name or brand, or who changes its intended purpose, may take on the obligations of a manufacturer. This is important for private-label arrangements and should be settled in writing before the first order.

A document checklist for a first shipment

ItemWhat to look for
EU declaration of conformityRefers to the MDR, identifies the manufacturer and the exact devices, states the class, and is signed and dated
Notified body certificate, for sterile devicesIssued to the manufacturer, covers the device group, is within validity; the body’s number matches the number beside the CE mark
Authorised representativeNamed on the label; a written mandate exists
Label and pack artworkManufacturer details, lot number, use-by date, sterility and single-use symbols, medical device symbol, UDI carrier, language requirements of the destination countries
Performance evidenceEN 13795 test reports stating part, edition and performance class for the finished product
Sterilisation and packagingValidation summaries for the sterilisation process and the sterile barrier system, with the stated shelf life
Quality systemCertificate with a scope covering the manufacture of the devices at the stated site

Common gaps to watch for

Several problems recur in practice. A declaration may refer to the earlier Medical Devices Directive; whether that is still acceptable depends on the transitional provisions, which have been amended more than once and need to be checked for the specific device. A certificate may be genuine but issued to a different legal entity or for a different product group. Labels may be correct in English but missing the languages required by the member state of sale. Customised variants ordered by the importer may not be covered by the manufacturer’s existing documentation. And a CE mark may appear on a product for which no notified body number is shown even though it is sold sterile.

Markets outside the Union that have historically followed CE marking, including the United Kingdom, now have their own rules, and these should be checked separately.

What to ask your supplier

Ask for the declaration of conformity and the artwork before you ask for a price. Ask who the authorised representative is, which notified body is involved for sterile products, and how a customised size or pack would be handled in the technical documentation. Agree who is the legal manufacturer for any private-label line. Shrisay Medical is an Indian manufacturer and exporter of surgical drapes, gowns and procedure packs; overseas buyers can raise market-specific regulatory questions through our export page or the contact page, and we will state plainly what documentation is available for a given product.

Frequently asked questions

Do sterile surgical drapes need a notified body under the MDR?

Class I devices supplied sterile require notified body involvement limited to the aspects relating to sterility. The CE mark on such products should be accompanied by the notified body’s identification number.

Is the importer responsible if the manufacturer’s paperwork is wrong?

Importers have their own obligations to verify key points before placing a device on the market and must not supply a device they believe is non-compliant. The detail of liability should be discussed with a regulatory or legal adviser.

Does private labelling change who the manufacturer is?

It can. An importer or distributor who places a device on the market under its own name may assume the obligations of a manufacturer, so the arrangement should be defined in a written agreement.